Regarding Oxidation-Reduction Potential (ORP) sensor
How should one select the electrode material for redox measurements: gold, silver, or platinum?
ORP Sensor Instruction Manual
Redox (ORP) Measurement Theory
What does a REDOX/ORP Sensor measure, and how does it measure it?
How should one select the electrode material for redox measurements: gold, silver, or platinum?
Welcome to leave a message for relevant information, we will contact you promptly!
How should one select the electrode material for redox measurements: gold, silver, or platinum?
Platinum is a versatile electrode material and is unsuitable only in very few applications.
In an ozone (O3) or hydrogen peroxide (H2O2) environment, platinum acts as a catalyst; therefore, at low concentrations, the potential corresponding to O3 or H2O2 cannot be established in the composite layer.
Furthermore, platinum oxidizes in air, let alone in strong oxidizing agents. In the former case, the electrode indicates an excessively high redox value, whereas in the latter case, a memory effect occurs. In both scenarios, the response is slow.
In principle, platinum electrodes should not be used in applications involving ozone or peroxides.
Gold electrodes can be used in the aforementioned applications; moreover, gold electrodes have a more positive potential than platinum electrodes. Gold or silver electrodes also offer advantages when employed in the treatment of cyanide-containing wastewater. Cyanide exerts a mild corrosive effect on the electrode surface, which helps keep the electrode surface clean.
Gold should not be used in acidic solutions with high chloride content, as gold forms complexes with chlorine, rendering the measurement invalid. The same principles that apply to pH measurement also apply to redox measurement, both in terms of instrument design and cable configuration: the measuring amplifier must have a very high input impedance (10¹² Ω). Redox values are typically displayed as absolute or relative “mV” readings.
Redox measurement is a bulk-parameter measurement and cannot be used to quantify a specific ion. Consequently, it is also impossible to determine the concentration of a particular species in a redox system, since all ions present in the solution participate in the redox reactions. Furthermore, redox electrodes should not be chemically cleaned in the medium being measured, because the cleaning solutions typically contain high concentrations of ions; such high ionic strength combined with the resulting potential shock can cause the electrode potential to return to its original value very slowly.
If cleaning is necessary during the process, it must be performed mechanically or by ultrasonic means.
Before use, the electrode can be briefly immersed in a sulfite solution, followed by treatment with sulfuric acid or hydrochloric acid. Failure of a redox measurement chain typically originates in the reference system, manifesting as zero-point drift. In such cases, zero-point calibration is required, for example using a 225 mV redox buffer solution. The slope of the redox electrode remains unchanged.
Next page:
Regarding pH
Regarding Conductivity
Regarding Oxidation-Reduction Potential (ORP) sensor
Regarding Dissolved Oxygen Sensor