As shown in the figure:
● Can be used with various analog conductivity Sensors
● Supports one thermoresistance input for automatic temperature compensation in conductivity measurements.
● Supports tag entry and temperature correction functions.
● Can be equipped with various types of conductivity, TDS, and salinity Sensors
● Provides a 4–20 mA output channel for transmitting the primary measured parameter or temperature measurement; RS485 output is also available as an option.
● Provides two relay outputs (upper/lower limit alarms)
● Provides password management permissions for operations to prevent unauthorized modification of parameter settings.
● Automatic temperature compensation
● Default factory setting is Chinese; supports both Chinese and English menu display.
● The entire unit features an IP67 protection rating, enabling outdoor, open-air installation and operation.
Technical Indicators:
Main Functions | Display/Transmit/Alarm/Communication |
Screen Specifications | 2.8-inch high-definition monochrome screen (320×240) |
Measurement range | 0–400 mS/cm |
Resolution | 0.01, 0.1, 1, 10 |
Accuracy | ±5%FS; ±0.3℃ |
Temperature compensation thermistor | NTC10K, NTC2.252K, PT100, PT1000 |
Temperature range | 0–99.9°C |
Operating Conditions | Ambient temperature: 0–60°C |
Input impedance | ≥1012 Ω |
Output | 4–20 mA (load < 750 Ω) |
Communication interface | RS485 |
Power supply | Compatible with AC 220 VAC and DC 24 V |
Communication interface | RS485 |
Protection rating | IP67 |
Relative humidity | <85% |
External dimensions | 113.5*118*136mm |
Installation size | 92*92mm |
Installation Method | Panel / Pipe / Wall Mounting |